Specification
A number of devices which 'crimp' a curved needle to a fine medical thread needed to be tested. The crimp force was applied horizontally pulling a lever on the device. A test system was required to verify that the force applied to the device was within the specified tolerance, and was repeatable over a number of tests.
Solution
The VersaTest motorised stand was adapted for horizontal operation by using a simple feet kit. The medical suture device was mounted on an 'L' bracket, with a dowel pin to secure it in position. An 'eye-loop' was connected to the force gauge on a clevis-pin and the connected to the lever device.The lever was then pulled at a test speed of 300mm/min with peak tension force being recorded on the AFG. The first data acquisition speed of 1200 Hz for the AFG, produced results more repeatable than those produced by the previous test system.
System
- 50 N AFG
- VersaTest Motorised Test Stand (superseded by the MultiTest-dV series)
- Horizontal Feet Kit
- 'L' Bracket and Flexible Eye-Loop'
Supplied to
Glaxo Wellcome, Poznan, Poland